 When using the mSupply ColdChain application for the first time, the Sensor Settings screen will have two sections: When using the mSupply ColdChain application for the first time, the Sensor Settings screen will have two sections:
   *Options   *Options
-  *Found Sensors +  *Found Sensors 
 +<WRAP center round important 60%> Please refer to the [[en:cold_chain:getting_started|mSupply ColdChain Getting Started]] for information about the device settings and sensor requirements and recommendations when setting up the application for the first time. </WRAP>
  
 === Options === === Options ===
-**Logging interval:** Sets the default number of minutes between logs on sensors. Using the slider, you can set the default log interval to any length between 1 and 100 minutes. The logging interval applies for all sensors when they are first linked to the mSupply ColdChain application. Different log intervals can be set for each sensor in **''Available Sensors/Log interval.''**+**Logging interval:** Sets the default number of minutes between logs on sensors. Using the slider, you can set the default log interval to any length between 1 and 100 minutes. The logging interval applies for all sensors when they are first linked to the mSupply ColdChain application. Different log intervals can be set for each sensor in **Available Sensors/Log interval** (see below).
 {{ :en:cold_chain:pasted:20210215-215500.png?300 }} {{ :en:cold_chain:pasted:20210215-215500.png?300 }}
  
Line 32: Line 33:
 {{ :en:cold_chain:pasted:20210215-215831.png }} {{ :en:cold_chain:pasted:20210215-215831.png }}
  
-<WRAP center round tip 30%>We recommend you avoid the use special characters (e.g. -, +, *) in the sensor name. </WRAP>+<WRAP center round tip 50%>We recommend you avoid the use special characters (e.g. -, +, *) in the sensor name. </WRAP>
  
  
Line 39: Line 40:
  
  
-<WRAP center round important 30%> We recommend that each device should be linked to a maximum of six sensors. Each sensor should only be linked to one device. </WRAP>  
  
 ==== Temperature breach settings ==== ==== Temperature breach settings ====
Line 56: Line 56:
 However, you can enter a custom name for each temperature breach category if required.  However, you can enter a custom name for each temperature breach category if required. 
 {{ :en:cold_chain:pasted:20210215-224931.png?350 }} {{ :en:cold_chain:pasted:20210215-224931.png?350 }}
-<WRAP center round tip 30%>We recommend you avoid the use special characters (e.g. -, +, *) when naming breach categories. </WRAP>+<WRAP center round tip 50%>We recommend you avoid the use special characters (e.g. -, +, *) when naming breach categories. </WRAP>
  
 **Duration:** Using the slider, you can determine how long the temperature must continuously remain outside of the required temperature range for it to be considered a breach. You can set the duration to any length between 1 and 600 minutes. **Duration:** Using the slider, you can determine how long the temperature must continuously remain outside of the required temperature range for it to be considered a breach. You can set the duration to any length between 1 and 600 minutes.
 {{ :en:cold_chain:pasted:20210215-225005.png }}  {{ :en:cold_chain:pasted:20210215-225005.png }}
-<WRAP center round important 30%>The duration should be equal or greater to the log interval. The duration should also be a multiple of the log interval to ensure breaches are captured accurately.</WRAP>+<WRAP center round important 50%>The duration should be equal or greater to the log interval. The duration should also be a multiple of the log interval to ensure breaches are captured accurately.</WRAP>
  
 **Temperature:** Using the slider, you can set the temperature outside of which will result in a breach: **Temperature:** Using the slider, you can set the temperature outside of which will result in a breach:
Line 77: Line 77:
 You can enter a name for each temperature breach category. You can enter a name for each temperature breach category.
 {{ :en:cold_chain:pasted:20210215-225113.png?350 }}  {{ :en:cold_chain:pasted:20210215-225113.png?350 }} 
-<WRAP center round tip 30%>We recommend you avoid the use special characters (e.g. -, +, *) when naming breach categories. </WRAP>+<WRAP center round tip 50%>We recommend you avoid the use special characters (e.g. -, +, *) when naming breach categories. </WRAP>
  
 **Duration:** Using the slider, you can determine the total time (sum) the temperature must remain outside of the required temperature range for it to be considered a cumulative breach. You can set the duration to any length between 1 and 600 minutes. **Duration:** Using the slider, you can determine the total time (sum) the temperature must remain outside of the required temperature range for it to be considered a cumulative breach. You can set the duration to any length between 1 and 600 minutes.
Line 88: Line 88:
 **Period:** The time period used for cumulative temperature configuration (and hence alerts) on the Sensor tab is set at 72 hours. **Period:** The time period used for cumulative temperature configuration (and hence alerts) on the Sensor tab is set at 72 hours.
  
-To view cumulative temperature breaches for a different period, you can **''change the time period on the Sensor Detail screen''**. The mSupply ColdChain application will use this time period to calculate cumulative temperature and display any breaches at the top of the Sensor Detail screen. +To view cumulative temperature breaches for a different period, you can [[en:cold_chain:sensors#sensor_detail|change the time period on the Sensor Detail screen]]. The mSupply ColdChain application will use this time period to calculate cumulative temperature and display any breaches at the top of the Sensor Detail screen.
- +
-<WRAP center round info 60%> You can export the sensor logs for further information and investigation regarding temperature breaches. </WRAP>+
  
 +<wrap center round info 95%> You can export the sensor logs for further information and investigation regarding temperature breaches. </wrap>
 +\\
 +\\
 +|  //  Previous:  **[[en:cold_chain:sensors]]** | | Return to homepage: **[[en:cold_chain:start]]** //  |
  
  • en/cold_chain/settings.1613429642.txt.gz
  • Last modified: 2021/02/15 22:54
  • by rowena